HP E1312A and HP E1412A Multimeter Specifications 179Appendix A
General Specifications
Overvoltage Category 1 (1500V peak max impulse)
HP E1312A and HP E1412A Available Power (Amps):
+5V: Ipm (maximum peak current): 0.20A
Idm (maximum dynamic current): 0.10A
+12V: Ipm (maximum peak current): 0.70A
Idm (maximum dynamic current): 0.06A
Cooling /Slot: Average Watts/Slot: 9.40
deltaP mm H2O: 0.05
Air Flow liters/s: 0.80
Operating Environment: 0
°
C to 55
°
C
65% Relative Humidity to 40
°
C
NOTE: Recalibration may be required after exposure to humidity levels >65%.
Storage Environment: -40
°
C to 70
°
C
State Storage Memory: Power-off state automatically saved
Warm-up Time: 1 hour
Programming Language: SCPI-1993, IEEE-488.2
The HP E1312A is not recommended for use in the HP E1300A or HP E1301A B-size Mainframe.