Emerson Process Management 3144P Marine Instruments User Manual


 
Reference Manual
00809-0100-4021, Rev DA
November 2004
4-29
Rosemount 3144P
FOUNDATION Fieldbus
LCD Display NOTE
The bar graph is not used for Foundation fieldbus devices, nor are the
“Sensor 1, Differential, Sensor 2, Multidrop, or Bust Mode” annunciators.”
The following are the default tags for each of the possible Function blocks
which display data on the LCD
Table 4-14. LCD Display Error Warning Descriptions
Message LCD top line LCD bottom line
RB.DETAILED_STATUS
Sensor Transducer Block Error “Error” “DVICE”
Manufacturing Block Integrity Error “Error” “DVICE”
Hardware/Software Incompatible “Error” “DVICE”
Non-volatile Memory Integrity Error “Error” “DVICE”
ROM Integrity Error “Error” “DVICE”
Lost Deferred NV Data “Error” “DVICE”
NV Writes Deferred No Errors Displayed
ADB Transducer Block Error No Errors Displayed
STB.SENSR_DETAILED_STATUS
Invalid Configuration “Error” “SNSOR”
ASIC RCV Error “Error” “SNSOR”
ASIC TX Error “Error” “SNSOR”
ASIC Interrupt Error “Error” “SNSOR”
Reference Error “Error” “SNSOR”
ASIC Configuration Error “Error” “SNSOR”
Sensor 1 Open “Error” “SNSOR”
Sensor 1 Shorted “Error” “SNSOR”
Terminal Temperature Failure “Error” “SNSOR”
Sensor 1 Out of Operating Range No Errors Displayed
Sensor 1 Beyond Operating Limits “Error” “SNSOR”
Terminal Temperature Out of Operating Range No Errors Displayed
Terminal Temperature Beyond Operating Limits “Error” “SNSOR”
Sensor 1 Degraded “Error” “SNSOR”
Calibration Error “Error” “SNSOR”
Sensor 2 Open “Error” “SNSOR”
Sensor 2 Shorted “Error” “SNSOR”
Sensor 2 Out of Operating Range No Errors Displayed
Sensor 2 Beyond Operating Limits “Error” “SNSOR”
Sensor 2 Degraded “Error” “SNSOR”
Sensor Drift Alarm “Error” “SNSOR”
Hot Backup Active “Error” “SNSOR”
Block Name LCD bottom line
Transducer “TRANS”
AI 1400 “AI 14”
AI 1500 “AI 15”
AI 1600 “AI 16”
PID 1700 “PID 1”
PID 1800 “PID 1”
ISEL 1900 “ISEL”
CHAR 2000 “CHAR”
ARITH 2100 “ARITH”
OSPL 2200 “OSPL”