Table 3-5. AMS Error Warning Descriptions
(1)
Message Description
Command not implemented The connected device does not support this function.
Communication error Either a device sends back a response indicating that the message it received was unintelligible,
or the 375 Field Communicator cannot understand the response from the device.
Device busy The connected device is busy performing another task.
Device disappears from list Device fails to respond to a command.
Device write protected Device is in write-protect mode. Data can not be written.
Illegal character An invalid character for the variable type was entered.
Illegal date The day portion of the date is invalid.
Illegal month The month portion of the date is invalid.
Illegal year The year portion of the date is invalid.
Incomplete exponent The exponent of a scientific notation floating point variable
is incomplete.
Incomplete field The value entered is not complete for the variable type.
Sending data to device Data is being transferred from configuration memory to a device.
There are write only variables which have
not been edited.
Please edit them
There are write-only variables that have not been set by the user. These variables should be set
or invalid values may be sent to the device.
There is unsent data. Send it before
shutting off?
Press YES to send unsent data and turn the 375 Field Communicator off. Press NO to turn the
375 Field Communicator off and lose the unsent data.
Too few data
bytes received
Command returns fewer data bytes than expected as determined by the device description.
Transmitter fault Device returns a command response indicating a fault with the connected device.
Units for <variable label> has changed.
Unit must be sent before editing, or invalid
data will be sent.
The engineering units for this variable have been edited. Send engineering units to the device
before editing this variable.
Unsent data to online device. SEND or
LOSE data
There is unsent data for a previously connected device which must be sent or thrown away
before connecting to another device.
Value out of range The user-entered value is either not within the range for the given type and size of variable or not
within the min/max specified by the device.
<message> occurred reading/writing
<variable label>
Either a read/write command indicates too few data bytes received, transmitter fault, invalid
response code, invalid response command, invalid reply data field, or failed pre- or post-read
method; or a response code of any class other than SUCCESS is returned reading a particular
variable.
<variable label> has an unknown value.
Unit must be sent before editing, or invalid
data will be sent.
A variable related to this variable has been edited. Send related variable to the device before
editing this variable.