Emerson Process Management 3144P Marine Instruments User Manual


 
Reference Manual
00809-0100-4021, Rev DA
November 2004
3-5
Rosemount 3144P
HART
AMS One of the key benefits of intelligent devices is the ease of device
configuration. When used with AMS, the 3144P is easy to configure and
provides instant and accurate alerts and alarms. The main configuration
screen of the 3144P is the “Configuration Properties” screen. From this
screen, the transmitter set-up can easily be viewed and edited.
The screens use a color-coding to give visual indication of the transmitter
health and to indicate any changes that may need to be made or written to the
transmitter.
Gray screens: indicates that all information has been written to the
transmitter
Yellow on screen: changes have been made in the software but not
sent to the transmitter
Green on screen: all current changes on screen have been written to
the transmitter
Red on screen: indicates an alarm or alert that requires immediate
investigation
Apply AMS Changes Changes made in the software must be sent to the transmitter in order for the
changes to take effect in the process.
1. From the bottom of the “Configuration Properties” screen, click Apply.
2. An “Apply Parameter Modification” screen appears, enter desired
information and click OK.
3. After carefully reading the warning provided, select OK.
REVIEW
CONFIGURATION DATA
Before operating the 3144P in an actual installation, review all of the
factory-set configuration data to ensure that it reflects the current application.
Review
Review the transmitter configuration parameters set at the factory to ensure
accuracy and compatibility with the particular application. After activating the
Review function, scroll through the data list to check each variable. If changes
to the transmitter configuration data are necessary, refer to “Configuration”
below.
AMS
Right click on the device and select “Configuration Properties” from the
menu. Select the tabs to review the transmitter configuration data.
HART Fast Keys 1, 4