Appendix F: Configuring Using HART
February 2015
Note
The display (reported) units of the transmitter’s process value (PV) can be changed to
metric or imperial measurement units using the parameter Primary Variable Units
(page 120). However, this does not automatically re-scale the PV. Use the parameter PV
Scale Factor (page 121) to manually re-scale the value.
F.2.2 Set as empt y
If the bottom reference is unknown and the tank is empty, the transmitter can change the
Bottom Reference value to the Distance measurement with the tank empty.
P010 = (D910 - P060)
Where:
P010 = Bottom Reference setting (see page 105).
D910 = Distance measurement (see page 153).
P060 = Distance Offset setting (see page 109).
Field Communicator
To select the Set As Empty command:
1. From the Home screen, select 2: Configure.
2. Select 2: Manual Setup.
3. Select 2: Level., and then 3: Present Depth.
4. Select 2: Set as Empty, and then press “ENTER”.
Note
When messages appear, take appropriate action if needed and press “OK”.
The command Set As Empty is also at Fast Key sequence 2, 1, 2.
Mobrey MCU900 Series control unit
To select the Set As Empty command:
1. From the Main Menu screen, select SETUP.
2. Select the transmitter (e.g. “Tx1:MSP400RH”).
3. Select DUTY, and then SET AS EMPTY.
4. Follow on-screen instructions to perform the Set As Empty action.
(If prompted to change the mode to off-line, press the Enter key).
5. Select “Quit” to exit to the previous menu.
Fast Keys
2, 2, 2, 3, 2
Select:
1. Present Depth
2. Set as Empty
3. Exit
2. Set as Empty
ENTER
ABORT
(Field Communicator Screen)
Esc=Quit =Start
SET AS EMPTY