Emerson Mobrey MSP400RH and MSP900GH Marine Radio User Manual


 
96
Reference Manual
IP2048/RM, Rev AA
Appendix F: Configuring Using HART
February 2015
F.2.7 Restar t device
This re-starts the transmitter as if the power has been interrupted.
Field Communicator
To restore the original factory configuration:
1. From the Home screen, select 3: Service Tools.
2. Select 4: Maintenance.
3. Select 3: Utilities, and then 1: Restart Device.
4. When the message “About to restart the transmitter”
appears, press “OK” to continue (or “ABORT” to not continue).
5. Select 1: Yes to restart the transmitter.
Note
When messages appear, take appropriate action if needed and press “OK”.
Mobrey MCU900 Series control unit
To restore the original factory configuration:
1. From the Main Menu screen, select SETUP.
2. Select the transmitter (e.g. “Tx1:MSP400RH”).
3. Select SYSTEM.
4. Select RESTART DEVICE.
5. Follow the on-screen instruction (“Start”) to restart the transmitter.
(Press the Enter ( ) key if prompted to change the mode to “off-line”).
6. Select “Quit” to exit to the previous menu.
Fast Keys
3, 4, 3, 1
Utilities
1. Restart Device
2. Load Defaults
3. Set Base Units
1. Restart Device
SAVE
HOME
HELP
(Field Communicator Screen)
Esc=Quit =Start
RESTART DEVICE
(Mobrey MCU901 Screen)