Section 4: Starting up
February 2015
4.2.6 Power Factor for the Flow Law
Screen display: P.FACt
Factory default value: 1.000
This menu is offered if the selected duty is Flow and a profile (e.g. “SPEC.C”) requires the
manual editing of a power factor in the formula:
Flow Q = kh* (where * = the power factor)
The transmitter is pre-programmed with appropriate power factors for many standard flow
profiles, and will automatically select the appropriate factor (See Table 4-1 on page 4-30).
Alternatively, the power factor may be edited to suit a specific flow structure.
Note
This menu option is in the programming menu. See Figure C-1 on page 76 for a map of
the menu structure and how to access the menu options.
To change the power factor
1. If entering the menu system from the PV display, press the green button repeatedly
until the “P.FACt” menu option is indicated (see note above.)
2. Press the blue button
to enter the “P.FACt” menu and display the present power
factor.
3. If the power factor is correct, press the red button
to exit to the menu. (The menu
option “P.FACt” re-appears. To get to the next menu, press

4. Press the blue button to start the editing mode. The first digit flashes to indicate it
can now be edited.
5. Press the green button
repeatedly to edit the flashing digit.
6. Press the blue button
to move to the next digit. The digit flashes to indicate it can be
edited.
7. Repeat steps (5) and (6) until the last digit is flashing, and edited as required.
8. Press the blue button
to confirm the new value. (The flashing stops).
9. If the new value is correct, press the red button
to save. The display changes to the
next menu option.
10. If the new value is incorrect, press the blue button
to exit to the menu.
11. The “P.FACt” menu re-appears to re-start at step (2) or press
for next menu.