162
Reference Manual
IP2048/RM, Rev AA
Appendix F: Configuring Using HART
February 2015
F.4.16 Transmit power (D919)
This indicates the level of power in use for ultrasonic pulse transmission.
The lower the level number, the less power is being used
The higher the level number, the more power being used
Note
The factory default setting is to optimize the power needed for ultrasonic pulse
transmission (see Transmit Power Control on page 135).
Field Communicator
To view the pulse transmission power in use:
1. From the Home screen, select 3: Service Tools.
2. Select 1: Alerts.
3. Select 2: Diagnostic Data.
4. Select 3: Operation.
5. Select 4: Transmit Power D919.
6. When finished, press EXIT to exit to the previous menu.
Note
This is also at the Fast Key sequence 2, 2, 7 [or 8], 4, 3
(see Appendix E: Field Communicator).
Mobrey MCU900 Series control unit
To view the pulse transmission power in use:
1. From the Main Menu screen, select MONITOR.
2. Select the transmitter (e.g. “Tx1:MSP400RH”).
3. Select DIAGNOSTICS.
4. Select OPERATION.
5. Select Transmit Power.
6. When finished, press the ESC key to exit to the previous menu.
Fast Keys
3, 1, 2, 3, 4
Transmit Power D919
9
EXIT
HELP
(Field Communicator Screen)
Transmit Power D919
9
(Mobrey MCU901 Screen)