161
Reference Manual
IP2048/RM, Rev AA
Appendix F: Configuring Using HART
February 2015
F.4.15 Pulses in use (D918)
This is the actual number of pulses transmitted in the previous burst of pulses.
Note
The factory default setting is to automatically decide the number of pulses in a burst for
optimum performance (see Target Pulses on page 139)
Field Communicator
To view the number of pulses being used in a burst:
1. From the Home screen, select 3: Service Tools.
2. Select 1: Alerts.
3. Select 2: Diagnostic Data.
4. Select 3: Operation.
5. Select 3: Pulses In Use D918.
6. When finished, press EXIT to exit to the previous menu.
Note
This is also at the Fast Key sequence 2, 2, 7 [or 8], 5, 2
(see Appendix E: Field Communicator).
Mobrey MCU900 Series control unit
To view the number of pulses being used in a burst:
1. From the Main Menu screen, select MONITOR.
2. Select the transmitter (e.g. “Tx1:MSP400RH”).
3. Select DIAGNOSTICS.
4. Select OPERATION.
5. Select Pulses In Use.
6. When finished, press the ESC key to exit to the previous menu.
Fast Keys
3, 1, 2, 3, 3
Pulses in Use D918
10
EXIT
HELP
(Field Communicator Screen)
Pulses in Use D918
10
(Mobrey MCU901 Screen)