To change the flow profile
1. If entering the menu system from the PV display, press the green button repeatedly
until the “ProF” menu option is indicated (see note on previous page).
2. Press the blue button
to enter the “ProF” menu and display the present profile
selection.
3. If the selected profile is correct, press the red button
to exit to the menu. (The menu
option “ProF” re-appears. To get to the next menu, press
4. Press the blue button to start the editing mode. The selected profile flashes to
indicate it can now be edited.
5. Press the green button
repeatedly to scroll through the list of profiles (see Table 4-1
on page 4-30.)
6. Press the blue button
to confirm the new profile. (The flashing stops.)
7. If the new profile is correct, press the red button
to save. The display changes to the
next menu option.
8. If the new profile is incorrect, press the blue button
to exit to the menu. The “ProF”
re-appears; re-start at step (2) or press
for the next menu.
Note
The next menu option will depend upon the flow profile chosen:
3/2 or 5/2: the transmitter will automatically calculate the Power factor and only
requires the K factor to be entered. (see “K-factor for the flow law” on page 32).
Manning: the next menu option is “LEUEL @ max”
(see “Maximum level entry” on page 33).
Parshall, FF, or FP: the transmitter will automatically calculate the appropriate Power
factor and K factor, and will set the 4 mA point at zero flow and the 20 mA point at
maximum flow. (See “Setting the output damping” on page 38).