Emerson Process Management 5300 Marine RADAR User Manual


 
Reference Manual
00809-0100-4530, Rev BA
July 2009
7-33
Rosemount 5300 Series
DIAGNOSTIC
MESSAGES
Troubleshooting If there is a malfunction despite the absence of diagnostic messages, see
Table 7-3 for information on possible causes.
Table 7-3. Troubleshooting
chart
Symptom Possible cause Action
No level reading
Power disconnected
Data communication cables
disconnected
Probe is not connected
Check the power supply
Check the cables for serial data
communication
View the Diagnostic window, see
“Diagnostics” on page 7-23 in order to
check active status messages
Check if “Probe Missing” is active. If it is,
check the probe connection
No HART communication.
COM Port configuration does not match
the connected COM Port
Cables may be disconnected
Wrong HART address is used
Hardware failure
•HART resistor
Check that correct COM Port is selected in
the HART server (see “Specifying the COM
Port” on page 5-15
Check wiring diagram
Verify that the 250 resistor is in the loop.
Check cables
Make sure that correct HART short address
is used. Try address=0
Check Analog Output current value to verify
that transmitter hardware works
Check that the correct settings are used in
RRM. Select Device, Search from the
menu. Click Settings and HART tab. Make
sure the values are properly chosen. The
standard values are shown on page 5-10
Analog Out is set in Alarm. Measurement or transmitter failure. View the Diagnostic window, see
“Diagnostics” on page 7-23 to check active
error and status messages.
Both Surface pulse and Interface
Pulse are detected, but Interface
Level is reported as unknown in the
Echo Curve plot.
Measurement Mode is set to “Level Only”. Set Measurement Mode to “Level and
Interface” (see “Basic Configuration
Parameters” on page 5-4).
Both Surface pulse and Interface
pulse are detected, but Interface
Level is reported as unknown in the
Echo Curve plot.
Interface pulse is identified as a double
bounce
Surface pulse and Interface pulse are
very close
No action required. Use the Echo Curve plot
to verify that the surface and interface are
close, see “Analyzing the Measurement
Signal” on page 7-3.
Surface pulse is detected, but Level is
incorrectly reported as Full or Empty.
Wrong Probe Type set
Bad Reference Threshold value
View the Diagnostics window, see
“Diagnostics” on page 7-23, to check active
messages and check if the warning “Full
Tank/Empty Tank” is active. If this is the case,
check that:
the transmitter is configured with correct
probe type
the reference pulse is below the reference
amplitude threshold. If not, adjust
reference threshold to an appropriate
value