Emerson Process Management 5300 Marine RADAR User Manual


 
Reference Manual
00809-0100-4530, Rev BA
July 2009
A-3
Rosemount 5300 Series
Mechanical (continued)
Dimensions See “Dimensional Drawings” on page A-9.
Probe Angle 0 to 90 degrees.
Housing / Enclosure Polyurethane-covered Aluminum or SST Grade CF8M (ASTM A743).
Flanges, Threads See “Ordering Information” on page A-21.
Height Above Flange See “Ordering Information” on page A-21.
Weight Transmitter Head (TH): 4.4 lbs (2 kg). SST 10.8 lbs (4.9 kg).
Flange: depends on flange size.
Coaxial probe: 0.67 lbs/ft (1 kg/m).
Rigid Single Lead probe (0.3 in./8 mm): 0.27 lbs/ft (0.4 kg/m).
Rigid Single Lead probe (0.5 in./13 mm): 0.71 lbs/ft (1.06 kg/m).
Rigid Twin Lead probe: 0.40 lbs/ft (0.6 kg/m).
Flexible Single Lead probe: 0.05 lbs/ft (0.08 kg/m).
Flexible Twin Lead probe: 0.09 lbs/ft (0.14 kg/m).
End weight: 0.88 lbs (0.40 kg) for the 4 mm single lead probe, 1.2 lbs (0.55 kg) for the 6 mm single lead probe
and 1.3 lbs (0.60 kg) for twin lead probes.
Environment
Ambient Temperature Non-Hazardous, HART communication: -40°F to 176°F (-40°C to 80°C).
IS/EEx ia and XP/EEx d, HART communication: -58°F to 158°F (-50°C to 70°C).
IS/EEx ia and XP/EEx d, F
OUNDATION Fieldbus: -58°F to 140°F (-50°C to 60°C).
LCD readable in: -4°F to 158°F (-20°C to 70°C).
Storage Temperature -58°F to 194°F (-50°C to 90°C). LCD: -40°F to 185°F (-40°C to 85°C).
Process Temperature
(7)
Standard: -40°F to +302°F ( -40°C to +150°C )
HTHP: -76°F to +752°F ( -60°C to +400°C )
HP: -76°F to +392°F ( -60°C to +200°C )
C: -320°F to +392°F (-196°C to +200 °C)
See temperature and pressure diagrams on page A-4.
Process Pressure
(7)
Standard: Full vacuum to 580 psig ( -1 to 40 Bar ).
HTHP: Full vacuum to 5000 psig (-1 to 345 Bar).
HP: Full vacuum to 5000 psig (-1 to 345 Bar).
C: Full vacuum to 5000 psig (-1 to 345 Bar).
See temperature and pressure diagrams on page A-4.
Humidity 0 - 100% Relative Humidity.
Ingress Protection NEMA 4X, IP 66 and IP67.
Telecommunication (FCC
and R&TTE)
FCC part 15 (1998) subpart B and R&TTE (EU directive 99/5/EC). Considered to be an unintentional radiator
under the Part 15 rules.
Factory Sealed Yes.
Vibration Resistance Aluminum housing: IEC 60770-1 Level 1. Stainless Steel housing: IACS E10.
Electromagnetic
Compatibility
Emission and Immunity: EMC directive 204/108/EC. EN61326-1: 2006. NAMUR recommendations NE21.
Built-in Lightning Protection EN61326, IEC 801-5, level 1 kV. T1 option: the transmitter complies with IEEE 587 Category B transient
protection and IEEE 472 surge protection
Pressure Equipment
Directive (PED)
Complies with 97/23/EC article 3.3.
(1) For probes with spacers, the accuracy may deviate close to the spacers.
(2) If the reference conditions are not met, an offset adjustment for the zero reference point may be necessary. The offset may be up to
± 25 mm.
(3) When using remote housing, the accuracy may be reduced.
(4) Not applicable for F
OUNDATION Fieldbus.
(5) PFA is a fluoropolymer with properties similar to PTFE.
(6) 1 mm PTFE cover.
(7) Final rating may be lower depending on flange and O-ring selection.