Emerson Process Management 5300 Marine RADAR User Manual


 
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
www.rosemount.com
Appendix A Reference Data
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-1
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-9
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-21
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-30
SPECIFICATIONS
General
Product Rosemount 5300 Series Guided Wave Radar Level and Interface Transmitter;
Model 5301, Liquid Level or Interface Transmitter (interface available for fully submerged probe).
Model 5302 Liquid Level and Interface Transmitter.
Model 5303 Solids Level Transmitter.
Measurement Principle Time Domain Reflectometry (TDR).
Reference Conditions Single Standard probe, 77°F (25°C) in water and ambient pressure.
Microwave Output Power Nominal 300 µW, Max. 45 mW.
CE-mark Complies with applicable directives (EMC, ATEX).
Safety Integrity Level 5300 FMEDA suitable for SIL2: The 5300 Series has been evaluated by Exida per hardware assessment
IEC61508. With a SFF > 90% it is Prior-use SIL2 suitable. For more information, go to:
http://www.emersonprocess.com/rosemount/safety/. Option code is QS.
Start-up Time < 40 s
Measuring Performance
Reference Accuracy ± 0.12 in. (3 mm) or 0.03% of measured distance, whichever is greatest.
(1)
(2)
(3)
Repeatability ± 0.04 in. (1 mm).
Ambient Temperature
Effect
± 0.008 in. (0.2 mm) /°K or ± 30 ppm/°K of measured value, whichever is greatest.
Update Interval < 1 per second.
Measuring Range 16 in. (0.4 m) to 164 ft (50 m).
Display / Configuration
Integral Display The integral digital display can toggle between: level, distance, volume, internal temperature, interface distance,
interface level, peak amplitudes, interface thickness, percentage of range, analog current out.
Note! The display cannot be used for configuration purposes.
Output Variables All models: Level, Distance to Level, Volume, Level Rate, Signal Strength, Internal Temperature, Signal Quality,
Surface/Noise Margin, Vapor DC, Analog Output Current
(4)
and % of Range
(4)
.
Model 5301 (in addition to the above for the case with fully submerged probe): Interface Level and Interface
Distance.
Model 5302 (in addition to the above): Interface Level, Interface Level Rate, Interface Distance, Upper Volume,
Lower Volume and Upper Product Thickness.
Output Units Level, Interface and Distance: ft, inches, m, cm or mm.
Level Rate: ft/s, m/s, in./min, m/h.
Volume: ft
3
, inch
3
, US gals, Imp gals, barrels, yd
3
, m
3
or liters.
Temperature: °F and °C.
Configuration Tools HART: Rosemount Radar Master, Rosemount 375 Field Communicator, AMS Suite or any other DD (Device
Description) compatible host system.
F
OUNDATION Fieldbus: Rosemount Radar Master, 375 Handheld Communicator, DeltaV
®
or any other DD
(Device Description) compatible host system.
FOUNDATION Fieldbus
Blocks
Resource block, 3 Transducer blocks, 6 AI blocks, PID block, ISEL block, SGCR block, ARTH block, and OS
block.