Rosemount 5300 Series
7-34
The reference pulse is not detected.
The tank is full
The transmitter is configured with
wrong probe type
Reference Amplitude Threshold is
incorrect
Check the product level
Check that correct probe type is configured
Check Reference Amplitude Threshold
Incorrect Interface Level reading.
Interface threshold incorrect
Upper Product Dielectric constant
incorrect
Adjust Interface threshold, see “Interface
Pulse not Found” on page 7-8
Check the Upper Product Dielectric
constant, see “Basic Configuration
Parameters” on page 5-4
Incorrect level reading.
Configuration error
Disturbing objects in the tank
Check the Tank Height parameter
Check status information and diagnostic
information
Check that the transmitter has not locked
on an interfering object, see “Disturbance
Echo Handling” on page 7-7
Adjust the Surface amplitude threshold, see
“Surface Pulse Not Found” on page 7-5
Integral display does not work.
Check the display configuration
Check loop power
Check Display connection
Contact Emerson Process Management
Service Department
(1)
FOUNDATION Fieldbus Card to
Transmitter Communication Fault
Verify Device Mode setting, should be
F
OUNDATION Fieldbus
(Parameter: ENV_DEVICE_MODE)
Restart method from Resource Block
Reboot gauge (Cycle Power)
Level Measurement Failure
Check Power Supply
Check the gauge configuration (Transducer
Block)
Check that the mechanical installation is
correct
Temperature Measurement Failure
Check ambient temperature
(2)
Restart gauge
Contact Emerson Process Management
Service Department
Volume Measurement Failure
Restart gauge
Check gauge configuration using PC Based
configuration tool
No surface echo
Check signal strength
Restart transmitter
See “Surface Pulse Not Found” on
page 7-5
DB Error/ Microwave Unit Error/
Configuration Error/ Other Error
Restart transmitter
Set database to default; load default
Database
Download Application Software
Call Service Center
SW Error/ Display Error/ Analog Out
Error
Restart transmitter
Call Service Center
(1) A malfunctioning display panel may only be replaced by service personnel at the Emerson Process Management Service Department. A display
must not be replaced when the transmitter is in operation.
(2) If the 5300 transmitter has been exposed to temperatures outside the specified limits, the device may stop its normal operation.
Symptom Possible cause Action