Tektronix CSA 803C Marine Radio User Manual


 
Performance Tests
Performance Verification
4Ć40
ăStep 5:ăMake the following selections:
AUTOSET button press.........................................
Slightly adjust the Trigger Level (in Trigger major menu), if necessary
MEASURE button press........................................
Measurement (in major menu) touch.........................
PeakĆPeak (in popĆup menu) touch...........................
WAVEFORM button press.......................................
Sampling Head Fnc's (in major menu) touch..................
Smoothing On (in popĆup menu) touch.......................
Exit (in popĆup menu) touch.................................
ăStep 6:ăCheck waveform for stable triggering. Using AUTOSET, repeat
this procedure for the following frequency generator settings:
400 mV
pĆp
at 1000 MHz
400 mV
pĆp
at 3.0 GHz