This procedure shows the setups and lists the steps to check the vertical
accuracy of the CSA 803C.
Setup to Check Vertical Gain
CSA 803C
Prec. Voltage Source
50 W Coaxial Cable
Calibration Head
DIRECT Input
Procedure to Check Vertical Gain
ăStep 1:ăInstall a Calibration Head (067-1413-00) into the position
shown in the setup diagram (CHĂ1/CHĂ2 slot). If the unit is in Standby
mode, set the power switch to ON. Set the Precision Voltage Source to
0 V and connect it to Channel 2 of the Calibration Head.
NOTE
If powering up from a cold start, the diagnostics will report an error
using the Calibration Head. The report will indicate a Time Base
error. To clear this, exit the diagnostics. Then, ignore the subseĆ
quent time base calibration failure notice and continue with the
following steps.
ăStep 2:ăInitialize the mainframe as follows:
UTILITY button press...........................................
Initialize (in major menu) touch..............................
Initialize (in verification popĆup menu) touch..................
ăStep 3:ăTrigger the CSA 803C as follows:
TRIGGER button press.........................................
Source (in major menu) touch...............................
Internal Clock (in popĆup menu) touch.......................
Vertical Accuracy