Performance Tests
Performance Verification
4Ć22
ăStep 5:ăSet the Voltmeter for DC Voltage, and check for a reading of
-5 V, ±200 mV.
If the readings noted in Steps 3 or 5 are outside the stated limits,
then servicing of the CSA 803C is required before continuing to the
Horizontal Reference Clock procedure.
GND+5 -5
Test Points
Figure 4Ć1:ăA26 M/F Acquisition Interconnect Board Test Point Locations