Tektronix CSA 803C Marine Radio User Manual


 
Performance Verification
Performance Verification
4Ć2
TableĂ4Ć1:ăMeasurement Limits, Specifications, and Functional Tests (Cont.)
Procedure Description Functional
Tests
Specifications
(Check)
Measurement
Limits (Examine)
Sweep Rate Accuracy 0.05% at 2 ns/div
0.08% at 1 ns/div
0.8% at 100 ps/div
2.5% at 10 ps/div
10% at 1 ps/div
none yes
Triggering yes
300 MHz Sensitivity 100 mV
pĆp
at 300 MHz 100 mV stable display
1000 MHz Sensitivity 100 mV
pĆp
at 1000 MHz 100 mV stable display
3.0 GHz Sensitivity 100 mV
pĆp
at 3.0 GHz 100 mV stable display
Internal Clock yes
Rise Time v2.5 ns v2.5 ns
Frequency
100 kHz ±3% 100 kHz ±3%
Duty Cycle 50% ±3% 50% ±3%
Prescaler 800 mV
pĆp
at 2 GHz
600 mV
pĆp
at 3 GHz
600 mV
pĆp
at 10 GHz
800 mV stable display
600 mV stable display
600 mV stable display
yes
Calibrator Output yes
Rise Time (typical) 250 ps 250 ps
PeakĆtoĆpeak Amplitude 250 mV ±10% 250 mV ±10%