Tektronix CSA 803C Marine Radio User Manual


 
Other Troubleshooting
Maintenance
6Ć104
See Figures 6Ć2 and 6Ć37 for the location of the line fuse and the A14 I/O
board fuses, respectively.
When a fuse must be replaced, unsolder the fuse from the board. Be careful
not to damage the solder pads on the board. (It may be helpful to straighten
the fuse leads on the rear of the board before removing the leads from their
holes in the circuit board.) Refer to the Parts List section for the correct value
and part number of each fuse.
TableĂ6Ć17:ăA14 I/O Board Fuse Failures
Fuse Executive Kernel Test Failure
(refer to Table 6Ć10, Executive
Subsystem Error Index Codes)
SelfĆTest/Extended Test Failure
F200
open
10hex (GPIB Interrupt)
F800
open
13hex (Front Panel Inter)
NOTE: The front panel lights do
not work so the code must be
read from the error status test
points (TP200ĆTP205) on the A17
Executive Processor board.
F600
open
Passes the Kernel diagnostic
tests, but the front panel lights
are not lit.
Exec Control E1411ă3
ąNVRAM E1411ă3
ąąBattery E1411ă1
ąă*Data Lines E1421ă1
ąă*Addr/Data E1431ă1
Internal I/O E3111ă1
ąTemp Sensor E3111ă1
ąąComparator
E3111ă1
ă*Tone Gen ĊĊ
ăą*Ramp Tone (works)
NOTE: The front panel lights,
soft keys, and hard keys do not
work.
F602
open
13hex (Front Panel Inter)
NOTE: The tone generator has a
very different tone.
*indicates a Manual Test forced by the operator. The test is not automatically executed by
SelfĆTest diagnostics.