Tektronix CSA 803C Marine Radio User Manual


 
Performance Tests
Performance Verification
4Ć46
ăStep 6:ăOn the CSA 803C, press the AUTOSET button.
ăStep 7:ăCheck the peakĆtoĆpeak amplitude as follows:
MEASURE button press........................................
Measurements (in the major menu) touch.....................
PeakĆPeak (in the popĆup menu) touch.......................
Rise (in the popĆup menu) touch.............................
Rise (in the major menu) touch..............................
Proximal (in the popĆup menu) touch.........................
Adjust Proximal for 20% and Distal for 80%.
Check that peakĆtoĆpeak amplitude is within 212.5 mV and 287.5 mV
(250 mV ±15%).
Check that rise time is approximately 250Ăps.