Step 9. Remove the flush short by holding its body and un
-
screwing its connector.
Step 10. Install the gauge onto the end of the sliding termi
-
nation.
Step 11. If the COARSE SET adjustment—which has been
set at the factory—has not moved, the inner dial on
the gauge will read “0.” If it doesn’t, perform the
Coarse Set Adjustment in Step 15.
Step 12. Place the sliding termination, with the gauge at
-
tached, on the bench top.
SLIDING TERMINATION MEASUREMENT CALIBRATION
7-16 37xxxE OM