Anritsu 37xxxE Marine Lighting User Manual


 
Step 7. When menu C3G appears, if you want to change
microstrip parameters to be different from those
shown in blue, place the cursor on MICROSTRIP
PARAMETERS and press the Enter key.
Step 8. When menu C16 (left) appears, move the cursor to
the Anritsu 3680 UTF calibration kit you wish to use
or to USER DEFINED; then press the Enter key.
The calibration kit selections shown in menu C16
are for the following 3680 Connection Substrate
Kits:
10 MIL KIT 36804B-10M
15 MIL KIT 36804B-15M
25 MIL KIT 36804B-25M
If you choose USER DEFINED, the next menu that
appears (C16A), lets you characterize your parame
-
ters. Move the cursor to each selection, key in a
value, then press the Enter key to return to menu
C16.
37xxxE OM 7-39
MEASUREMENT CALIBRATION LRL/LRM CALIBRATION
MENU C3G
CONFIRM
CALIBRATION
PARAMETERS
LRL/LRM
PARAMETERS
MICROSTRIP
PARAMETERS
USER DEFINED
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE
MENU C16
SELECT
MICROSTRIP
KIT TO USE
10 MIL KIT
15 MIL KIT
25 MIL KIT
USER DEFINED
PRESS <ENTER>
WHEN COMPLETE
MENU C16A
ENTER
MICROSTRIP
PARAMETERS
WIDTH OF
STRIP
XX.XXXX mm
THICKNESS OF
SUBSTRATE
XXXX.XXXX mm
Zc
XXX.XXX pW
SUBSTRATE
DIELECTRIC
XX.XX
EFFECTIVE
DIELECTRIC
XX.XX
(RECOMMENDED
0.00)
PRESS <ENTER>