A-216 37xxxE OM
S ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
SET SCALING
OR PRESS
<AUTOSCALE>
None
–REAL– None
RESOLUTION
XX.XXX U/DIV
Sets the resolution for the vertical axis of the active
channel’s displayed graph. Resolution can by set
incrementally using the keypad or rotary knob.
SCL; SCL?;
SCL2; SCL2?
REFERENCE VALUE
XXX.XXX pU
Sets the value at the reference line for the active channel
amplitude measurement on the displayed graph. The value
can be set in increments of 0.001 U using the keypad or
rotary knob.
OFF; OFF?
OFF2; OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel’s amplitude
measurement. This is the line about which the amplitude
expands with different resolution values. The reference line
can be set to any vertical division using the rotary knob.
REF; REF?;
REF2; REF2?
PRESS <ENTER>
TO RESUME CAL
On the CAL_SS10 menu, pressing the Enter key returns
you to the calibration setup or sequence.
None
Menu SS10 or CAL_SS10, Set Scaling 10