MENU DESCRIPTION GPIB COMMAND
CALIBRATE FOR
FLAT PORT POWER
None
FORWARD
DIRECTION ONLY
None
XXX POINTS
MEASURE 1 PWR
POINT EVERY
XX POINT(S)
Displays the number of power points (0 50) to be skipped
during the power sweep. The points not measured are
interpolated to provide a flat sweep.
PTS; PTS?
POWER TARGET
-XXX.X dBm
Lets users set a flat output-power value (power target). The
VNA defaults to Port 1 power.
PTP; PTP?
START FLAT
POWER CALIBRATION
Begins the calibration. If calibration is successful, you are
returned to menu SU8. If the calibration unsuccessful due
to a fatal error (Source or power meter inoperable or not
connected), this menus remains displayed. At any time,
you can abort the calibration by pressing the DEFAULT
PROGRAM or CLEAR/RET LOC keys. All other keys are
locked out.
SFC
PRESS <ENTER>
TO SELECT
TURN KNOB TO
CHANGE NUMBER
OF POINTS
Pressing the Enter key implements your menu selection or
turns the function on/off.
None
Menu SU8 or CAL_SU8, Calibrate For Flat Test Port Power