Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-19
ALPHABETICAL LISTING A
MENU DESCRIPTION GPIB COMMAND
AUTOCAL
CHARACTERIZATION
None
SWITCH AVERAGING
XXXX
Enter an appropriate amount of SWITCH AVERAGING
(recommend 4 for the electronic modules, and 16 for the
electromechanical modules).
ACSW; ACSW?
PORT CONFIG
L=1, R=2
R=1, L=2
Switch between the side of the AutoCal module which is
connected to Port 1 and Port 2 (default LEFT connected to
Port 1, RIGHT connected to Port 2).
ACL1R2;
ACR1L2;
ACARP?
NUMBER OF AVGS None
REFLECTION
XXXX
Enter the number of averages to be used with the reflection
standards in the AutoCal module (default 10 averages).
ACRFL; ACRFL?
LOAD
XXXX
Enter the number of averages to be used with the load
standard in the AutoCal module (default 10 averages).
ACLO; ACLO?
THRU
XXXX
Enter the number of averages to be used with the thru
standard (default 4 averages).
ACTUAVG;
ACTUAVG?
ISOLATION
XXXX
Enter the number of averages to be used with the isolation
standard in the AutoCal module (default 32 averages).
ACISO; ACISO?
START AUTOCAL
CHARACTERIZATION
Calls menu CAL_SEQ, which starts the AutoCal
characterization sequencing.
None
PRESS <ENTER>
TO SELECT
OR SWITCH
Press the Enter key to select or switch. None
Menu ACAL_CHAR, AutoCal Characterization Menu