37xxxE OM A-67
ALPHABETICAL LISTING C
MENU DESCRIPTION GPIB COMMAND
CHANGE LRL/LRM
PARAMETERS
None
NEXT CAL STEP Moves to the next calibration step. Must move cursor to
here after making below selections. Pressing the Enter key
then moves you to the next step.
None
REFLECTION
OFFSET LENGTH
+XXX.XXXX mm
Enter the offset length of the reflective device. None
REFLECTION TYPE ROL; ROL?
GREATER THAN Zo Specifies the reflection to have an impedance value greater
than the reference impedance (Z
0).
This is typically an open
device.
RGZ; RXZ?
LESS THAN Zo Selects the reflection to have an impedance value less than
the reference impedance (Z
0).
This is typically a short
device.
RLZ; RXZ
MATCH PARAMETERS
MATCH IMPEDANCE
+XXX.XXX W
Allows entry of the impedance (defaults to 50.000 W). LMZ; LMZ?
MATCH INDUCTANCE
+XXXX.XXXX pH
Allows entry of the inductance (defaults to 0.00 e-12). LMZL; LMZL?
PRESS <ENTER>
TO SELECT
Pressing the Enter key implements the selection. None
Menu C19, Change LRL/LRM Parameters