Step 3.
Select CHANGE CAL METHOD AND LINE
TYPE, in menu C11 (left). (This assumes LRL and
MICROSTRIP are not presently shown in blue as be
-
ing selected.)
Step 4. When menu C11A (bottom left) appears, highlight
the following selections.
a.
LRL/LRM and press the Enter key.
b.
MICROSTRIP and press the Enter key.
c.
NEXT CAL STEP and press the Enter key.
Step 5. When menu C11 reappears, confirm that the
LRL/LRM calibration method and MICROSTRIP
line-type have been selected. Select NEXT CAL
STEP and press the Enter key to proceed.
Step 6. Continue through the calibration sequence, and
make the following selections from the menus that
appear:
INCLUDE ISOLATION (STANDARD) (Menu
C5D)
NORMAL (1601 POINTS MAXIMUM) (Menu C1)
START (Your start frequency) (Menu C2)
STOP (Your stop frequency) (Menu C2)
LRL/LRM CALIBRATION MEASUREMENT CALIBRATION
7-38 37xxxE OM
MENU C11
BEGIN CALIBRATION
KEEP EXISTING
CAL DATA
REPEAT
PREVIOUS CAL
AUTOCAL
CAL METHOD
XXXXXXXX
TRANSMISSION
LINE TYPE:
XXXXXXXX
CHANGE CAL
METHOD AND
LINE TYPE
NEXT CAL STEP
PRESS <ENTER>
TO SELECT
MENU C11A
CHANGE CAL METHOD
AND LINE TYPE
NEXT CAL STEP
CAL METHOD
SOLT (STANDARD)
SSLT (DOUBLE OFF
-
SET SHORT WITH
LOAD)
SSST (TRIPLE OFF
-
SET SHORT)
LRL/LRM
TRM
TRANSMISSION
LINE TYPE
COAXIAL
WAVEGUIDE
MICROSTRIP
PRESS <ENTER>
TO SELECT
MENU C5D
SELECT USE
OF ISOLATION
IN CALIBRATION
INCLUDE
ISOLATION
(STANDARD)
EXCLUDE
ISOLATION
PRESS <ENTER>
TO SELECT
MENU C1
SELECT
CALIBRATION
DATA POINTS
NORMAL
(1601 POINTS
MAXIMUM)
C.W.
(1 POINT)
N-DISCRETE
FREQUENCIES
(2 TO 1601
POINTS)
TIME DOMAIN
(HARMONIC)
PRESS <ENTER>
TO SELECT
MENU C2
FREQ RANGE OF
CALIBRATION
START
0.0400000000GHz
STOP
20.000000000 GHz
201 DATA PTS
0.099800000 GHz
STEP SIZE
MAXIMUM NUMBER
OF DATA POINTS
1601 MAX PTS
801 MAX PTS
401 MAX PTS
201 MAX PTS
101 MAX PTS
51 MAX PTS
NEXT CAL STEP
PRESS <ENTER>
TO SELECT