Step 9. When menu C3 (left) appears, if you want to change
any of the parameters shown in blue letters, place
the cursor on that parameter and press the Enter
key. For this example, we will change them all, start
-
ing with the top one. Move the cursor to PORT 1
CONN and press the Enter key.
Step 10. In menu C4 (below), which appears next, move the
cursor to K CONN (M) and then press the Enter key.
This choice presumes that you have a K-Female con
-
nector on the device-under-test (DUT). Remember, in
this menu you choose the connector type on the test
port, or the connector type that mates with the DUT
connector. When menu C3 returns, observe that K
CONN (M) is now shown in blue for the PORT 1
CONN choice.
SOLT CALIBRATION MEASUREMENT CALIBRATION
7-22 37xxxE OM
MENU C3
CONFIRM
CALIBRATION
PARAMETERS
PORT 1 CONN
K CONN (M)
PORT 2 CONN
SMA (M)
REFLECTION
PAIRING
MIXED
LOAD TYPE
SLIDING
THROUGHLINE
PARAMETERS
REFERENCE
IMPEDANCE
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE
MENU C4
SELECT PORT 1
CONNECTOR TYPE
K-CONN (M)
K-CONN (F)
V-CONN (M)
V-CONN (F)
W1-CONN (M)
W1-CONN (F)
SMA (M)
SMA (F)
GPC-3.5 (M)
GPC-3.5 (F)
GPC-7
USER DEFINED
MORE
PRESS <ENTER>
TO SELECT