A-116 37xxxE OM
G ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
SWEPT POWER
GAIN COMPRESSION
None
PORT 1 ATTN
0*10 dB (0 - 70)
Attenuates the microwave source power at port 1 from 0 to
70 dB, in 10 dB steps. The power is attenuated before
being applied to Port 1 for a forward transmission or
reflection test (S
21
or S
11
, respectively).
SA1; SA1?
PORT 2 ATTN
0*10 dB (0 - 40)
Attenuates from 0 to 40 dB (10 dB steps) the microwave
power being input to Port 2 from the device-under-test
(DUT).
TA; TA2?
PREVIOUS MENU Returns to the previous menu. None
PRESS <ENTER>
TO SELECT
Pressing the Enter key implements your menu selection. None
Menu GC_SU2, Swept Power Gain Compression 2