Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-223
ALPHABETICAL LISTING S
MENU DESCRIPTION GPIB COMMAND
TEST SIGNALS None
POWER CONTROL
+XX.X dB
0 TO -15.0 dB)
Enter the delta-power level for the Port 1 output in dB. PWR; PWR?
PORT 1 ATTN
XX10 dB (0 - 70)
Attenuates the microwave source power at port 1 from 0 to
70 dB, in 10 dB steps. The power is attenuated before
being applied to Port 1 for a forward transmission or
reflection test (S
21
or S
11
, respectively).
SA1; SA1?
PORT 1 POWER
-XX.XX dBM
Displays the Port 1 power, in dBm. PIP?
PORT 2 ATTN
XX10 dB (0-X0)
Attenuates from 0 to 40 dB (10 dB steps) the microwave
power being input to Port 2 from the device-under-test
(DUT).
TA2; TA2?
CALIBRATE
FOR FLATNESS
(CAL EXISTS)
None
FLATNESS ON(OFF)
CORRECTION
AT XX.X dBm
Calls menu SU8 or CAL_SU8, depending on whether valid
Flat Test Port Power calibration data exists. Both of these
menus provide selection control for the Flat Test Port
Power feature.
FP0; FP1; FPX?
EXT SOURCE 1 PWR
-XX.XX dBm
Enter the power level, in dBm, of the 1st, external
frequency source.
PWR1; PWR1?
EXT SOURCE 2 PWR
-XX.XX dBm
Enter the power level, in dBm, of the 2nd, external
frequency source.
PW2; PW2?
PREVIOUS MENU Returns to the previous menu. None
Menu SU2 or CAL_SU2, Sweep Setup 2