Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-53
ALPHABETICAL LISTING C
MENU DESCRIPTION GPIB COMMAND
CHANGE CAL METHOD
AND LINE TYPE
None
NEXT CAL STEP Select next calibration step. Must move cursor to here after
making below selections. Pressing the Enter key then
moves you to the next step.
None
CAL METHOD None
(SOLT) STANDARD This option and the ones below allow you to select the
method (procedure) to be used to calibrate. This method is
independent of the calibration type, which may be 12-term,
reflection only, etc.
SCM; CMX?
SSLT (DOUBLE
OFFSET SHORT
WITH LOAD)
Selects the double offset short calibration method. OCM; CMX?
SSST (TRIPLE
OFFSET SHORT)
Selects the triple offset short calibration method. OC3M; CMX?
LRL/LRM Selects LRL or LRM method. LCM; CMX?
TRM Selects TRM method. TCM; CMX?
TRANSMISSION
LINE TYPE
None
COAXIAL Selects coaxial cable as the transmission line type. LTC; LTX?
WAVEGUIDE Selects waveguide as the transmission line type. LTW; LTX?
MICROSTRIP Selects microstrip as the transmission line type. LTU; LTX?
PRESS <ENTER>
TO SELECT
Pressing the Enter key implements your menu selection. None
Menu C11A, Select Calibration Method