Anritsu 37xxxE Marine Lighting User Manual


 
Operating Procedure, Receiver Mode
The three operational modes that comprise the Receiver Mode can be
set up as follows:
Step 1. Press the Option Menu key (below).
Step 2.
When menu OPTNS (top left) appears, select RE-
CEIVER MODE.
Step 3. When menu RCV1 (middle left) appears, select ei-
ther STANDARD (Step 4) or USER DEFINED
(Step 5). Your selection depends on the application.
Step 4. The Standard mode uses the Source Lock mode for
operation with the internal source. The user has no
control over selections within the Standard Mode.
Because entering the standard mode from the User
Defined Mode erases the current stored calibration
data, a warning menu (RCV3, bottom left) appears
when STANDARD is selected. Press Enter to enter
into the Standard mode or press Clear to abort.
NOTE
Spur Reduction: Normal/OFF: Spur Reduction Off may be
selected when making non-ratioed measurements or using
the Set-On Receiver mode. Under those measurement con
-
ditions, it may reduce high level noise. In normal
S-parameter measurement mode, Spur Reduction should
remain “Normal,” as the noise level is not affected.
RECEIVER MODE MEASUREMENTS
8-60 37xxxE OM
.
-
Measurement
Enhancement
Channels
Display
Option
Menu
MENU RCV3
STANDARD
RECEIVER MODE
WARNING:
CONTINUING
MAY INVALIDATE
CURRENT
SETUP AND
CALIBRATION
PRESS <ENTER>
TO CONTINUE
PRESS <CLEAR>
TO ABORT
MENU RCV1
RECEIVER MODE
STANDARD
USER DEFINED
SOURCE CONFIG
SPUR REDUCTION
NORMAL/OFF
PRESS <ENTER>
TO SELECT
MENU OPTNS
OPTIONS
TRIGGERS
REAR PANEL
OUTPUT
DIAGNOSTICS
MULTIPLE
SOURCE
CONTROL
RECEIVER MODE
SOURCE CONFIG
RF ON/OFF
DURING RETRACE
PRESS <ENTER>
TO SELECT