Step 13. When menu C3 returns:
a.
Observe that REFLECTION PARING now re
-
flects MIXED.
b.
Move cursor to LOAD TYPE and press the En
-
ter key. This brings up menu C6 (below).
This menu lets you select either of two load
types, broadband or sliding. Broadband loads
are adequate for all but the most demanding re-
flection measurements. They are easier to use
and less expensive than sliding loads. If you
choose a sliding load, refer to paragraph 7-3 for
a procedure on setting pin depth.
For this example, select BROADBAND LOAD
and press the Enter key.
c. The next menu to appear, C6A (left), prompts
you to enter an impedance value. For this exam
-
ple, use the rotary knob to change the displayed
value to 50W. Alternatively, you can key in
50 ohms. That is, press 50 on the keypad and
the X1 terminator key. If the value is 1 mW, key
in .001 and press the 10
–3
terminator key. Con
-
versely, if the value is 1 MW, key in 1000 and
press the 10
3
terminator key.
SOLT CALIBRATION MEASUREMENT CALIBRATION
7-24 37xxxE OM
MENU C6
SELECT
TYPE OF LOAD
BROADBAND
FIXED LOAD
SLIDING LOAD
(MAY ALSO
REQUIRE
BROADBAND
FIXED LOAD)
PRESS <ENTER>
TO SELECT
MENU C3
CONFIRM
CALIBRATION
PARAMETERS
PORT 1 CONN
TYPE N (M)
PORT 2 CONN
TYPE N (F)
REFLECTION
PARING
MIXED
LOAD TYPE
BROADBAND
THROUGH
PARAMETERS
REFERENCE
IMPEDANCE
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
MENU C6A
ENTER
BROADBAND LOAD
IMPEDANCE
BROADBAND
LOAD
IMPEDANCE
50.000 W
PRESS <ENTER>
TO SELECT