A-108 37xxxE OM
G ALPHABETICAL LISTING
- SWEPT POWER GAIN COMPRESSION -
- CALIBRATION INSTRUCTIONS -
1. SET FREQUENCIES (FROM 1 TO 10 FREQS ALLOWED).
2. P START POWER SHOULD BE APPROXIMATELY =
AUT (x dB compression spec) - AUT (gain) - 15 dB.
3. P STOP SHOULD BE 20 dB HIGHER THAN P START.
4. PORT 2 INPUT POWER SHOULD BE LESS THAN 0 dB
(UNLESS OPTION 6 IS INSTALLED).
5. INDICATE THE GAIN COMPRESSION POINT VALUE (x dB).
6. NOMINAL OFFSET = APPROXIMATE GAIN (OR LOSS)
OF EXTERNAL DEVICES PRECEDING THE AUT.
Menu EXT_GC2, Gain Compression Help Menu 2
PORT1 PORT2
AMP ATTN TEST PORTS ATTN
12
NOMINAL
OFFSET