37xxxE OM A-27
ALPHABETICAL LISTING C
MENU DESCRIPTION GPIB COMMAND
CAL FREQ RANGE None
CENTER
XXX.XXXXXXXXXGHz
Enter the center frequency for calibration. If you desire, you
can change this frequency for your measurement when you
reach menu SU1_CENTER, which follows the final
calibration menu.
CNTR; CNTR?
SPAN
XXX.XXXXXXXXX GHz
Enter the span width for calibration. Like the start
frequency, this too can be changed for your measurement.
SPAN; SPAN?
SET START/STOP Calls menu C2, which lets you enter a start and stop
frequency.
SRT; STP
XXX DATA POINTS
XXX.XXXXXXXXXGHz
STEPSIZE
The program automatically sets the step size, based on the
selected center and span frequencies. The step size will be
the smallest possible (largest number of points up to a
maximum of 1601), based on the chosen frequency span.
ONP
MAXIMUM NUMBER
OF DATA POINT(S)
1601 MAX PTS
801 MAX PTS
401 MAX PTS
201 MAX PTS
101 MAX PTS
51 MAX PTS
NP1601
NP801
NP401
NP201
NP101
NP51
NEXT CAL STEP Displays the next menu in the calibration sequence. None
PRESS <ENTER>
TO SELECT
Pressing the ENTER key implements your menu selection. None
Menu C2_CENTER, Frequency Range of Calibration (Center/Span)