Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-75
ALPHABETICAL LISTING C
MENU DESCRIPTION GPIB COMMAND
ADAPTER REMOVAL None
12-TERM CALS FOR
X AND Y MUST
EXIST IN THE
CURRENT DIRECTORY
None
ELECTRICAL LENGTH
OF THE ADAPTER
+XXX.XXXX ps
Displays the electrical length of the adapter. The value of
the electrical length is used when the two calibrations are
merged. It has the same range as the time delay for
reference plane extension and a default value of 0.0000 ps.
ADPL; ADPL?
REMOVE ADAPTER Calls menu CAR2, which leads you through the reading of
the Y’-Y and X-X’ calibration files and the computation of
the new 12-term error coefficients.
None
HELP Calls menu EXT_CAR, which provides help text for using
this feature.
None
PRESS <ENTER>
TO SELECT
Pressing the Enter key implements the selection. None
Menu CAR1, Adapter Removal 1
MENU DESCRIPTION GPIB COMMAND
ADAPTER REMOVAL None
READ CAL FILE OF
X TEST PORT
FROM SD CARD
(ADAPTER ON
PORT 2)
Calls menu DSK2, which provides instructions. None
READ CAL FILE OF
X TEST PORT
FROM USB DRIVE
(ADAPTER ON
PORT 2)
Calls menu DSK2, which provides instructions. None
PRESS <ENTER>
TO SELECT
Pressing the Enter key implements the selection. None
PRESS <CLEAR>
TO ABORT
Pressing the Clear key aborts the selection. None
Menu CAR2, Adapter Removal 2