Anritsu 37xxxE Marine Lighting User Manual


 
Chapter 5
Error and Status Messages
5-1 INTRODUCTION This chapter lists, describes, and provides corrective action for the er
-
ror messages that point to problems that the operator can correct. Any
error messages that appear on the display but do not appear in this
chapter will require action by a qualified service representative.
5-2 ERROR MESSAGES Error messages are provided in Tables 5-1 and 5-2.
37xxxE OM 5-3
Error Message Description Corrective Action
ATTENUATOR UNAVAILABLE Option 6 Port 2 Test Step Attenuator is
not installed.
Install Option 6 Step Attenuator,
BANDS MUST SEQUENCE Frequency bands in Multiple Source
mode must sequence in a 1-2-3-4-5 or-
der.
None, no skipping is allowed.
BOTH LIMITS MUST BE ON Must have both limits activated. Turn on limits.
DIFFERENT H/W SETUP. RECALL
ABORTED
Source is different from the recalled
setup.
Reconfigure system to duplicate the
hardware setup that was used to store
the saved data.
DIFFERENT S/W VERSION, RECALL
ABORTED
Saved state not compatible with hard
-
ware or software version.
Load compatible software (S/W) version
and retry.
DISCRETE FREQS LOST Change in frequency caused discrete fill
frequencies to be lost.
None.
DISPERSIVE MEDIUM, ONLY TIME
USED
Distance does not apply for dispersive
media.
None.
FREQUENCIES HAVE REACHED UPPER
LIMIT
Frequencies being defined in Multiple
Source mode have reached upper limits
of Sources.
Redefine frequencies to not exceed lim
-
its of Sources.
ILLEGAL IN C.W. MODE Attempted to readout limit frequency. None, no limit lines are permitted in CW
mode.
ILLEGAL IN TIME DOMAIN Attempted to readout limit frequency None.
LOGO FILE NOT FOUND Attempted to read a non-existent logo
file from disk.
Create user-defined logo using applica
-
tion on external controller.
MEAS DATA NOT AVAILABLE FOR
STORAGE
Measurement data is not available for
storage on the USB drive or SD Card.
None.
MEMORY LOCATION CORRUPTED Requested memory location is cor
-
rupted.
None. If problem reoccurs after storing a
new setup, contact Anritsu Customer
Service.
NO BANDS ARE STORED No frequency bands have been defined
and stored.
Need to define and store frequency
bands to turn on Multiple Source mode.
Table 5-1. General Error Messages (1 of 3)