Anritsu 37xxxE Marine Lighting User Manual


 
Step 11. With menu C3 (left) displayed, move the cursor to
PORT 2 CONN and press the Enter key. Following
the procedure in Step 10, select K CONN (M) for the
Port 2 connector.
Step 12. When menu C3 returns:
a.
Observe that PORT 2 CONN now reflects K
CONN (M).
b.
Move the cursor to REFLECTION PARING
and press the Enter key. This brings up menu
C13 (below).
Reflection Pairing lets you mix or match the
Open and Short reflection devices in the Cali
-
bration Sequence menus. The MIXED choice
lets you calibrate using first an Open on one
port and a Short on the other, then a Short on
one port and an Open on the other. Conversely,
MATCHED lets you calibrate first using an
Open on both ports then using a Short on both
ports. For this example, choose MIXED and
press the Enter key.
37xxxE OM 7-23
MEASUREMENT CALIBRATION SOLT CALIBRATION
MENU C3
CONFIRM
CALIBRATION
PARAMETERS
PORT 1 CONN
K CONN (M)
PORT 2 CONN
K CONN (M)
REFLECTION
PAIRING
MIXED
LOAD TYPE
SLIDING
THROUGH
PARAMETERS
REFERENCE
IMPEDANCE
TEST SIGNALS
START CAL
PRESS <ENTER>
TO SELECT
OR CHANGE
MENU C13
SELECT
REFLECTION
PAIRING
MIXED
(OPEN–SHORT
SHORT–OPEN)
MATCHED
(OPEN–OPEN
SHORT–SHORT)
PRESS <ENTER>
TO SELECT