Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-211
ALPHABETICAL LISTING S
MENU DESCRIPTION GPIB COMMAND
SET SCALING
OR PRESS
<AUTOSCALE>
None
–PHASE– None
RESOLUTION
XX.XXX °/DIV
Sets the resolution for the vertical axis of the active
channel’s displayed graph. Resolution can by set
incrementally using the keypad or rotary knob.
SCL; SCL?;
SCL2; SCL2?
REFERENCE VALUE
XXX.XXX °
Sets the value at the reference line for the active channel
amplitude measurement on the displayed graph. The value
can be set in increments of 0.01
o
using the keypad or
rotary knob.
OFF; OFF?
OFF2; OFF2?
REFERENCE LINE
X
Sets the reference line for the active channel’s amplitude
measurement. This is the line about which the amplitude
expands with different resolution values. The reference line
can be set to any vertical division using the rotary knob.
REF; REF?;
REF2; REF2?
PHASE SHIFT
X.XX °
Sets the value by which the active channel’s phase
measurement is shifted on the phase graph. The shift can
be set in increments of 0.01 degrees using the keypad or
rotary knob. This is useful when phase data is near the 180
degree rollover value.
PHO; PHO?
PRESS <ENTER>
TO RESUME CAL
On the CAL_SS5 menu, pressing the Enter key returns you
to the calibration setup or sequence.
v
Menu SS5 or CAL_SS5, Set Scaling 5