Step 7. Menu C1 appears next (left) and lets you select the
number of frequency points for which calibration
data is to be taken. Select NORMAL (1601 POINTS
MAXIMUM) (refer to section 7-4 for a description).
Step 8. The next menu, C2 (below), lets you set your start
and stop frequencies. For this example, move the cur
-
sor to START, press 65 on the keypad, then press
the GHz terminator key. Perform like operations for
the STOP choice, except make the entry read
110 GHz. After setting the frequencies, select NEXT
CAL STEP and press the Enter key.
Step 9. This brings up the Menu C14 (lower left) for select
-
ing the connector types on ports 1 and 2. Select the
W1-CONN connectors with the appropriate sex for a
W1 (1mm) calibration.
TRIPLE OFFSET-SHORT CALIBRATION (SSST) MEASUREMENT
CALIBRATION
7-34 37xxxE OM
MENU C2
FREQ RANGE OF
CALIBRATION
START
0.0400000000GHz
STOP
20.000000000 GHz
201 DATA PTS
0.099800000 GHz
STEP SIZE
MAXIMUM NUMBER
OF DATA POINTS
1601 MAX PTS
801 MAX PTS
401 MAX PTS
201 MAX PTS
101 MAX PTS
51 MAX PTS
NEXT CAL STEP
PRESS <ENTER>
TO SELECT
MENU C1
SELECT
CALIBRATION
DATA POINTS
NORMAL
(1601 POINTS
MAXIMUM)
C.W.
(1 POINT)
N-DISCRETE
FREQUENCIES
(2 TO 1601
POINTS)
TIME DOMAIN
(HARMONIC)
PRESS <ENTER>
TO SELECT
Menu C14
SELECT PORT n
OFFSET SHORT
CONNECTOR TYPE
W1-CONN (M)
W1-CONN (F)
SPECIAL A (M)
SPECIAL A (F)
SPECIAL B (M)
SPECIAL B (F)
SPECIAL C (M)
SPECIAL C (F)
USER DEFINED
PRESS <ENTER>
TO SELECT