A-118 37xxxE OM
G ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
LINEAR POWER
CALIBRATION
None
CALIBRATING FOR
LINEAR POWER...
None
PRESS <CLEAR>
TO ABORT
Pressing the Clear key aborts the Linear Power Calibration. None
Menu GC_SU8A-ABORT, Abort Calibrate for Linear Power
- LINEAR POWER CALIBRATION -
- CALIBRATION INSTRUCTIONS -
LINEAR POWER CALIBRATION ADJUSTS THE SOURCE OUTPUT
POWER FOR EACH POWER FREQUENCY POINT ACROSS THE POWER
SWEEP RANGE TO PROVIDE A LINEAR POWER LEVEL AT THE
TEST PORT (FORWARD DIRECTION ONLY).
- INSTRUCTIONS -
1. PRESET, ZERO, AND CALIBRATE THE POWER METER.
2. CREATE AND ACTIVATE THE POWER METER'S CAL
FACTOR LIST FOR THE POWER SENSOR BEING USED.
3. CONNECT THE POWER METER TO THE DEDICATED GPIB
INTERFACE AND THE POWER SENSOR TO THE TEST PORT.
4. SELECT <START LINEAR POWER CALIBRATION>.
Menu EXT_GC_SU8A, Gain Compression Help Menu