Anritsu 37xxxE Marine Lighting User Manual


 
LRL/LRM Calibration Setup Flow—Description
1. Pressing the Begin Cal key calls Menu C15.
2. With one exception, the flow is from left to right in the direction of
the arrow head. The exception occurs in Menu C1, for the TIME
DOMAIN choice. Here the flow direction reverses to Menu C2C
then returns to a left-to-right flow on to Menu C3E, C3G, or C3F.
3. Arrowheads that point both left and right indicate that the flow
returns to the right-most menu after a choice had been made.
4. The group of menus to the left of Menu C3E, C3G, or C3F are the
initial selection set and are essentially the same for all four
calibration types: Standard, Offset-Short, TRM, and LRL/LRM.
5. The group of menus that follow Menu C3E, C3G, or C3F are, for
the most part, type specific. The selection of Menu C3E, C3G, or
C3F depends upon the choice made in Menu C11A: COAXIAL,
WAVEGUIDE, or MICROSTRIP.