If you are creating a new file, enter the filename and
select DONE when finished.
10-7 USING AUTOCAL
An example procedure for using the AutoCal module is given below.
This example assumes a frequency range of 40 MHz to 40 GHz, a
power level of -7 dBm, and use of a Series 36582 AutoCal module.
Step 1. Press the Begin Cal key (top left).
Step 2.
Select AUTOCAL from the displayed menu (C11,
left).
37xxxE OM 10-9
AUTOCAL USING AUTOCAL
.
-
Measurement
Enhancement
Channels
Display
Begin
Cal
Apply
Cal
MENU C11
BEGIN CALIBRATION
KEEP EXISTING
CAL DATA
REPEAT
PREVIOUS CAL
AUTOCAL
CAL METHOD
XXXXXXXXX
TRANSMISSION
LINE TYPE:
XXXXXXXXX
CHANGE CAL
METHOD AND
LINE TYPE
NEXT CAL STEP
PRESS <ENTER>
TO SELECT