Anritsu 37xxxE Marine Lighting User Manual


 
A-32 37xxxE OM
C ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
CONFIRM
CALIBRATION
PARAMETERS
Used for Offset-Short method and coaxial line type. None
PORT 1 CONN
W1-CONN (M)
Calls menu C14 to select a connector offset short kit for
Port1.
None
PORT 2 CONN
W1-CONN (M)
Calls menu C14 to select a connector offset short kit for
Port2.
None
REFLECTION
PAIRING
XXXXXXXX
Calls menu C13A for a SSLT calibration or calls
menu C13B for a SSST calibration, which lets you select
the pairing (mixed or matched) for the types of reflection
devices (open/short) that you will use on Ports 1 and 2 for
calibration.
None
LOAD TYPE
XXXXXXXX
Calls menu C6, which displays type of load selected for
calibration—broadband, fixed, or sliding. Move cursor here
and press Enter to display the menu used to change the
load type.
None
THROUGH LINE
PARAMETERS
Calls menu C20, which lets you enter throughline
parameters—including offset length and loss equation
coefficients.
None
REFERENCE
IMPEDANCE
Calls menu C15, which lets you choose the reference
impedance value (1 mW to1MW) for the devices connected
to Ports 1 and 2 for calibration. Default value is 50W.
None
TEST SIGNALS Calls menu CAL_SU2, which lets you enter calibrate Flat
Test Port Power or change source power(s) and attenuator
settings.
None
START CAL Starts the offset-short calibration sequence using coaxial
standards.
BEG
PRESS <ENTER>
TO SELECT
OR CHANGE
Pressing the Enter key implements your menu selection. None
Menu C3A, Confirm Calibration Parameters