Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-123
ALPHABETICAL LISTING G
MENU DESCRIPTION GPIB COMMAND
SELECT
GRAPH TYPE
None
LOG MAGNITUDE Selects a log magnitude graph for display on the active
channel's selected S-parameter. The active channel is
indicated by its key (CH1, CH2, CH3, CH4) being lit.
MAG
PHASE Selects a phase graph for display on the active channel. PHA
LOG MAGNITUDE
AND PHASE
Selects log magnitude and phase graphs for display on the
active channel.
MPH
SMITH CHART
(IMPEDANCE)
Selects a Smith chart for display on the active channel. SMI; SME; SMC
SWR Selects an SWR display for the active channel. SWR
GROUP DELAY Selects a Group Delay display for the active channel. DLA
POWER OUT Provides for measuring output power. The measurement of
output power is accomplished by using the b2 (or Tb)
measured value normalized to the power supplied to the
AUT at Test Port 1. While the b2 parameter is the most
meaningful for this graph type, you may use any other
parameter.
POW
MORE Calls additional graph type selections on menu GT2. None
PRESS <ENTER>
TO SELECT
AND RESUME CAL
Pressing the Enter key implements your menu selection
(and resumes the calibration from where it left off, if in the
calibration mode).
None
Menu GT1/CAL_GT1, Select Graph Type