Step 5. Press the Begin Cal key (below).
Step 6. Follow the menu prompts and choose to perform a
full 12-term calibration. Use the Adapter’s SMA
male connector (Y') as Test Port 1 and the VNA’s Test
Port 2 connector as Test Port Y.
Step 7. Press the Save/Recall Menu key (below).
Step 8.
Choose SAVE from the displayed menu (top left).
Step 9. Choose the appropriate SD Card or USB drive loca
-
tion, based on individual preference (Menu SR2, bot
-
tom left).
Step 10. When prompted, select CREATE NEW FILE and en
-
ter a conventional DOS filename, such as YP
-
RIME_Y.CAL. (Store this file in the current direc
-
tory.)
Step 11. Now connect the Adapter’s SMA male end to the
VNA’s Test Port 2 SMA female connector.
ADAPTER REMOVAL MEASUREMENTS
8-36 37xxxE OM
.
-