A-26 37xxxE OM
C ALPHABETICAL LISTING
MENU DESCRIPTION GPIB COMMAND
FREQUENCY RANGE
OF CALIBRATION
None
START
XXX.XXXXXXXXXGHz
Enter the sweep-start frequency for calibration. If you
desire, you can change this frequency for your
measurement when you reach menu SU1, which follows
the final calibration menu. The only restriction is that your
start measurement frequency be greater than or equal to
your start calibration frequency.
SRT; SRT?
STOP
XXX.XXXXXXXXX GHz
Enter the sweep-stop frequency for calibration. Like the
start frequency, this too can be changed for your
measurement. The stop frequency must be lower than or
equal to your stop calibration frequency. In other words,
your measurement frequency span must be equal to or
smaller than your calibration frequency span.
STP; STP?
SET CENTER/SPAN Calls menu C2_CENTER, which lets you enter a center
frequency and span range.
CNTR; SPAN
XXX DATA POINTS
XXX.XXXXXXXXXGHz
STEPSIZE
The program automatically sets the step size, based on the
selected start and stop frequencies. The step size will be
the smallest possible (largest number of points up to a
maximum of 1601), based on the chosen frequency span.
ONP
MAXIMUM NUMBER
OF DATA POINT(S)
1601 MAX PTS
801 MAX PTS
401 MAX PTS
201 MAX PTS
101 MAX PTS
51 MAX PTS
NP1601
NP801
NP401
NP201
NP101
NP51
NEXT CAL STEP Displays the next menu in the calibration sequence. None
PRESS <ENTER>
TO SELECT
Pressing the ENTER key implements your menu selection. None
Menu C2, Frequency Range of Calibration (Start/Stop)