Anritsu 37xxxE Marine Lighting User Manual


 
37xxxE OM A-31
ALPHABETICAL LISTING C
MENU DESCRIPTION GPIB COMMAND
CONFIRM
CALIBRATION
PARAMETERS
Used for SOST method and coaxial line type. None
PORT 1 CONN
XXXXXXXX
Calls menu C4_P1 or C4A_P1, which displays the Port 1
test port connector type to be used during OSL calibration.
This should agree with the connector type that both your
calibration components and the test device mate with.
Move cursor here and press Enter to display menu used to
change connector type.
None
PORT 2 CONN
XXXXXXXX
Calls menu C4_P2 or C4A_P2, which displays the Port 2
test port connector type to be used during OSL calibration.
This should agree with the connector type that both your
calibration components and the test device mate with.
Move cursor here and press Enter to display menu used to
change connector type.
None
REFLECTION
PAIRING
XXXXXX
Calls menu C13, which lets you select the pairing (mixed or
matched) for the types of reflection devices (open/short)
that you will use on Ports 1 and 2 for calibration.
None
LOAD TYPE
XXXXXXXX
Calls menu C6, which displays type of load selected for
calibration—broadband fixed or sliding. Move cursor here
and press ENTER to display menu used to change load
type.
None
THROUGH LINE
PARAMETERS
Calls menu C20, which lets you enter throughline
parameters—including offset length and loss equation
coefficients.
None
REFERENCE
IMPEDANCE
Calls menu C17, which lets you choose the reference
impedance value (1 mW to1kW) for the devices connected
to Ports 1 and 2 for calibration. Default value is 50W.
None
TEST SIGNALS Calls menu CAL_SU2, which lets you enter calibrate Flat
Test Port Power or change source power(s) and attenuator
settings.
None
START CAL Starts the standard (OSL) calibration sequence using
coaxial standards.
BEG
PRESS <ENTER>
TO SELECT
OR CHANGE
Pressing the Enter key implements your menu selection. None
Menu C3, Confirm Calibration Parameters